TABLE 1: Air volume measurements
(means ± SD and mean errors) using three pneumotachometer settings
of the Aerosport KB1-C ambulatory metabolic analyzer (ten tests each) in
reference to volumes of air injected with a calibration syringe.
Calibration |
Low Flow |
|
Medium Flow |
|
High Flow |
|
(L/min) |
mean ± SD |
mean error |
mean ± SD |
mean error |
mean ± SD |
mean error |
3
2.63 ± 0.22 (12.3%)
1.89 ± 0.37 (37.0%)
0.15 ± 0.05 (95.0%)
6
5.07 ± 0.08 (15.5%)
5.82 ± 0.15 (3.0%)
0.65 ± 0.29 (89.2%)
9
8.64 ± 0.07 (4.0%)
8.84 ± 0.08 (1.8%)
2.87 ± 0.57 (68.1%)
18
17.37 ± 0.07 (3.5%)
17.47 ± 0.09 (2.9%)
15.46 ± 0.44 (14.1%)
27
26.31 " 0.41 (2.8%)
27.12 ± 0.55 (1.5%)
24.61 ± 0.21 (8.8%)
36
34.83 " 0.27 (3.2%)
36.94 ± 0.47 (2.6%)
33.06 ± 0.23 (8.2%)
45
43.88 " 0.08 (2.5%)
45.38 ± 0.78 (1.5%)
41.62 ± 0.22 (7.5%)
54
48.33 " 3.71 (10.5%)
53.53 ± 0.32 (0.9%)
50.24 ± 0.41 (7.0%)
63
50.81 " 0.74 (19.3%)
62.40 ± 0.27 (1.0%)
59.44 ± 0.37 (5.6%)
72
71.61 ± 0.21 (0.5%)
69.22 ± 0.29 (3.9%)
81
80.48 ± 0.24 (0.6%)
78.44 ± 0.26 (3.2%)
90
89.90 ± 0.60 (0.5%)
88.01 ± 0.46 (2.2%)
99
99.79 ± 0.41 (0.8%)
98.54 ± 0.31 (0.5%)
108
106.81 ± 1.10 (1.1%)
106.72 ± 0.40 (1.2%)
117
111.65 ± 1.51 (4.6%)
115.59 ± 0.41 (1.2%)
126
116.52 ± 1.49 (7.5%)
124.01 ± 0.36 (1.6%)
135
133.90 ± 0.60 (0.8%)
144
141.84 ± 0.95 (1.5%)
153
149.50 ± 0.34 (2.3%)
162
160.76 ± 0.57 (0.8%)
171
169.91 ± 0.56 (0.6%)
180
179.54 ± 2.83 (1.3%)
189
186.80 ± 1.68 (1.2%)
198
194.36 ± 1.58 (1.8%)
207
200.63 ± 1.58 (3.1%)
216
208.76 ± 2.78 (3.3%)